Configuration
69
2106778-001
7
Configuration
The configuration of the patient monitor consists of platform configuration and clinical
configuration.
This chapter describes how to perform the platform configuration tasks needed to take the
patient monitor into use for the first time and the configuration tasks needed for administration
and maintenance thereafter.
For information on how to perform the clinical configuration, including care unit settings and
user profiles, refer to the patient monitor’s user’s manual.
7.1 Adjusting display
7.1.1 Adjusting the display brightness
NOTE: Refer to the the patient monitor’s user’s manual for information how to setup the
screens:
Monitor Setup > Default Setup > Care Unit Settings > Screens
.
If needed, adjust the picture on the display using the Auto Adjustment feature in the display's
OSD menu. Refer to the display's user documentation for details.
7.1.2 Calibrating a touchscreen
Calibrate the connected touchscreens, one by one, as follows:
1.
Select
Monitor Setup > Service Calibrations.
2.
In the
Enter Password
dialog box, type the username and password and press
Enter
.
Username:
biomed
Password:
Change
<space>
Me
NOTE: Username and password are case sensitive.
NOTE: The factory default password for the username “biomed” is
“
Change Me”.
3.
In the
Service / Calibrations
menu, select
Touch Screen
.
4.
The Touch calibration screen opens.
5.
Touch the calibration mark or cross hair (+) in each corner of the screen, according to the
instructions shown on the screen.
NOTE: When you touch the cross hair in the upper left hand corner of the touch screen you are
calibrating, the other connected displays will turn blank until the calibration is completed.
7.2 Configuring the network
7.2.1 Configuring hostname
1.
Log in to Webmin.
2.
Select
Configuration >
Network > Hostname.
The
Hostname Configuration
window opens.
The current hostname is shown in the
Current Value
column.
3.
Enter the new hostname in the
Change Value to
column.
NOTE: The hostname is a unique, 4 to 32 character long identifier of a patient monitor in
the network. Use alphanumeric characters A-Z, a-z, 0-9. The hostname may include also
characters “-” and “_”, but it can't start or end with these characters.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......