CARESCAPE Monitor B850
102
2106778-001
•
Make sure that all the target peripheral devices are connected to the patient monitor.
NOTE: Do not disconnect any peripheral device, or shut down the patient monitor, until the
software activation is successfully completed.
To activate EMBC and UIC software:
1.
EMBC/UIC Software
window, in
Current Software State
, verify that the software you
want to activate is listed as
Inactive
.
The correct UIC software depends on the display REF number:
M1177405 is for D15K, D19KT with REF part number 2039143-001.
2064128-001 is for D19KT with part number 2067141-001 and for D19KT VER01.
2095419-001 for D19KT with REF part number 2042925-026 and USB Remote
controller.
If needed, transfer the correct UIC software before the activation.
2.
Select the peripheral devices whose software you want to activate.
3.
Select
Activate
to start the software activation to the selected peripheral devices.
4.
Confirm activation by selecting
Yes
in the Host Software window that opens.
NOTE: Software activation may take up to 10 minutes.
5.
According to your Webmin access:
•
If you are using Webmin on a service PC:
a.
EMBC Software – Wait until the software activation completes and the E-Module
Frame (F5/F7) performs an automatic restart. The green On/Standby (Power) LED in
the F5/F7 frame turns off when the EMBC software activation is started and
illuminates again after the EMBC software activation is completed.
b.
UIC Software - Wait until the software activation completes and the patient monitor
performs an automatic restart. It may take up to 10 minutes.
•
If you are using Webmin locally via the integrated browser of the patient monitor and you
are activating:
a.
EMBC Software - Wait until the software activation completes and the E-Module
Frame (F5/F7) performs an automatic restart. The green On/Standby (Power) LED in
the F5/F7 frame turns off when the EMBC software activation is started and
illuminates again after the EMBC software activation is completed.
b.
UIC Software - Select
Refresh
to start the activation, then select
Refresh
repeatedly
until the software activation progress bar shows 100%. Then wait until the software
activation completes and the patient monitor performs an automatic restart.
6.
Verify that the software activation is successful and the peripheral devices run the
activated software.
7.20.3 Canceling pending host software activation
You can cancel a pending host software activation at any time while the message
Software
activation after next discharge/after next case end
is shown on the display.
1.
Log in to Webmin.
2.
Select
Configuration
>
Software Management.
The
Software Management
window opens.
3.
Select
Host Software
from the
Software List
and select
Next
.
4.
In the
Host Software
window, below
Status
, verify that the
Software activation pending
is
Yes
.
5.
Select
Cancel
.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......