The secondary display
The secondary display must be a digital display. It can be either an USB or serial touchscreen,
or a non-touchscreen display. You cannot connect a cloned display to DVI-D 2 connector.
1.
Connect one end of the video cable to the DVI-D 2 connector in the rear of the patient
monitor. Connect the other end of the video cable to the display.
2.
If applicable, connect a USB or serial touchscreen cable to the USB 2 or RS-232 2
connector in the rear of the patient monitor. Connect the other end of the touchscreen
cable to the display.
The third display
The optional third display can only be used to support the iPanel application. The patient
monitor must be equipped with an optional third video card.
The third display can be either an analog or digital display. It can be either an USB touchscreen,
or a non-touchscreen display.
You may optionally connect a cloned analog display to the DVI-I 3 port. Use a DVI-I to DVI-D
and VGA splitter to do so. The patient monitor does not support touchscreen interface for the
cloned third display.
1.
Connect one end of the video cable to the DVI-I 3 connector in the rear of the patient
monitor. Connect the other end of the video cable to the display.
2.
If applicable, connect a USB touchscreen cable to the USB 3 connector in the rear of the
patient monitor. Connect the other end of the USB touchscreen cable to the display.
6.3.2 Connecting remote displays
NOTE: The instructions in this chapter applies to D19KT VER01 displays.
Display port 1
Display
Touchscreen interface
port
DVI-I 1
Display 1
USB1 or RS-232 1
Display 1 cloned
USB1 or RS-232 1
Display port 2
Display
Touchscreen interface
port
DVI-D 2
Display 2
USB2 or RS-232 2
Display port 3
Display
Touchscreen interface
port
DVI-I 3
Display 3
USB3
Display 3 cloned
No touchscreen interface
available
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......