Disassembly and reassembly
157
2106778-001
12.2.3 Detaching the uDOM
1.
Open the unit as described in
.
2.
Remove the screw that holds the uDOM to the processor board and detach the uDOM
from the processor board connector. Torque [0.3 Nm].
3.
Replace the uDOM.
4.
Reassemble in reverse order.
NOTE: uDOM, disk on module, is the permanent memory of the patient monitor. The software,
the licenses, and both clinical and service configurations are all stored to the uDOM.
Replacing the uDOM
Contact GE Service if you doubt that the uDOM is defective and needs to be replaced. Only GE
Service can configure the new uDOM.
When you contact GE, provide the following information:
a) Monitor Type
b) Serial Number
c) Software version installed, if available
12.2.4 Replacing the front bezel
Disassemble first the following:
a.
.
1.
Disconnect the light pipe pulling gently from the receptacle on the PCB.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......