Configuration
101
2106778-001
The software activation starts automatically after the clinical user performs a patient
discharge / case end the next time. The patient monitor shows a screen saver that informs
about the ongoing software activation:
Software activation in progress. Do not disconnect any measurement modules or other
peripheral devices, or shut down the monitor until the software activation is complete.
Activation may take up to 10 minutes. The device will automatically restart once the
software activation is complete.
NOTE: When the software is activated first time after a uDOM replacement, the screen saver
does not appear.
The clinical user must wait until the software activation is complete and the patient monitor
restarts automatically. If the patient monitor starts up normally and no error messages appear
on the display, the activation is successful.
Activating module software
Follow this instruction when you want to:
•
activate new sGAS software to a CARESCAPE Respiratory Module
•
activate new PDM software to a CARESCAPE Patient Data Module
•
activate new PiCCO software to an E-PiCCO Module
•
activate new NMT software to an E-NMT Module (only possible for E-NMT-01 version)
•
activate new LoFlo software to an E-LoFlo Module
Before you start:
•
Make sure that the patient monitor is in a case reset / patient discharge state.
•
Make sure that the target parameter module is connected to the patient monitor.
NOTE: Do not disconnect the parameter module, or shut down the patient monitor until the
software activation is completed and the parameter module has restarted.
NOTE: Do not activate the host software while the module software upgrade is in progress.
To activate module software:
1.
In the
PDM/sGAS/PiCCO/NMT/LoFlo Software
window, in
Current Software State
, verify
that the software you want to activate is listed as
Inactive
.
2.
Select
Activate
to activate the inactive software.
3.
Confirm activation by selecting
Yes
in the
Host Software
window that opens.
The patient monitor will now start activating the module software. It will show a message
PDM module removed
or
xxx measurement removed
(xxx refers to parameter name) on
the patient monitor screen. This message will remain until the module software activation
is completed and the parameter module has restarted. This may take up to 15 minutes.
Do not shut down the patient monitor or disconnect the parameter modules. The
parameter module restarts automatically after the software activation is complete.
4.
After the parameter module restarts, verify that the software activation is successful and
the parameter module runs the activated software.
Activating EMBC or UIC software
Follow this instruction when you want to:
•
activate new EMBC software to a F5 or F7 Frame
•
activate new UIC software to a display or USB Remote Controller
You can activate new software to several peripheral devices at a time.
Before you start:
•
Make sure that the patient monitor is in a case reset / patient discharged state.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......