CARESCAPE Monitor B850
B - 2 (4)
2106778-001
Single Fault Condition (SFC)
Polarity: Normal
1 mA
Polarity: Reversed
1 mA
Normal Condition (NC)
Polarity: Reversed
500 µA
EN /IEC
300 µA
UL
Normal Condition (NC)
Polarity: Normal
100 µA
Single Fault Condition (SFC)
Ground closed
Polarity: Normal
500 µA
EN /IEC
300 µA
UL
Ground open
Polarity: Normal
500 µA
EN /IEC
300 µA
UL
Ground open
Polarity: Reversed
500 µA
EN /IEC
300 µA
UL
Ground closed
Polarity: Reversed
500 µA
EN /IEC
300 µA
UL
Normal Condition (NC)
Polarity: Reversed
100 µA
Electrical Safety Test completion
Observed result
Acceptance criteria
PASS
N.A.
FAIL
Monitor starts up normally.
: Picture quality
Text is readable, images are clear and
brightness is good.
: Touchscreen control
Touchscreen is correctly calibrated
and operates correctly.
Hemodynamic parameter data
appears on the screen correctly.
E-module parameter data appears on
the screen correctly.
Hard key
Selected menu opens and selected
activity starts.
Trim Knob
Selected menu opens and selected
activity starts.
Selected menu opens and selected
activity starts.
Keyboard language configuration is
correct.
Data is correctly populated.
Network symbol and other monitor’s
patient data is correctly shown on the
screen.
Test page is printed to the selected
printer.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......