CARESCAPE Monitor B850
A - 2 (4)
2106778-001
Yes:
Yes: you do not have to perform the electrical safety tests. To continue the
installation check, mark the electrical safety test completion N.A. and
proceed to
. If available, attach the original electrical safety
test results to this check form.
No:
No: continue to perform the electrical safety tests.
Power outlet is correctly wired.
Power cord and plug are undamaged and all conductors are properly connected.
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
a.) Ground continuity test
without power cord
0.1 ohms
with power cord
0.2 ohms
b.) Impedance of protective earth
connection
without power cord
0.1 ohms
with power cord
0.2 ohms
Normal Condition (NC)
500 µA
EN /IEC
300 µA
UL
Single Fault Condition (SFC)
1 mA
Normal Condition (NC)
100 µA
Single Fault Condition (SFC)
500 µA
EN /IEC
300 µA
UL
Module type
S/N
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
ECG
SpO2/
InvP
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Polarity: NORMAL
10 µA
Single Fault Condition (SFC)
Ground closed (normal)
50 µA
Ground open (normal)
50 µA
Ground closed (reverse)
50 µA
Ground open (reverse)
50 µA
Normal Condition (NC)
Polarity: Reverse
10 µA
Patient (sink) leakage current tests
Polarity: NORMAL
50
µA
Polarity: Reverse
50
µA
Module type
S/N
Observed result
Acceptance
criteria
PASS
N.A.
FAIL
Patient (source) leakage current tests, using a test body
Normal Condition (NC)
Polarity: NORMAL
10 µA
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......