Configuration
71
2106778-001
•
CF:xx:xx:xx:xx:xx: Reserved by IANA for PPP
•
FF:FF:FF:FF:FF:FF: Broadcast
•
Multicast addresses in which the first octet is an odd number (for example,
97:xx:xx:xx:xx:xx)
9.
Select the applicable
Speed & Duplex
option.
10. Select
Save
.
The default network speed is 10 Mbps half duplex, which suppresses the effects of network
storms resulting from network misconfigurations. Please ensure that the network in your
facility meets the network configuration requirements as described in the Patient Monitoring
Network Configuration Guide.
The network configurations will be saved and become active when the patient monitor is
restarted.
7.2.3 Selecting and configuring S/5 Network
To select S/5 Network for the real-time network infrastructure:
1.
Log in to Webmin.
2.
Select
Configuration >
Network > Wired Interfaces
.
The current network configuration is shown in the
Present Configuration
table.
3.
Below
Select Network Type
in the
Network Configuration
window, select the applicable
network type:
S/5
.
4.
Select
Next
to proceed on network configuration.
NOTE: When
S/5
Network
is selected, the CARESCAPE Network is disabled.
The patient monitor may be configured to the S/5 Network, or IX Network, or both.
1.
In the
Network Configuration
window below
S/5 Network
, enter a
Virtual ID.
NOTE: Valid values are within the range of 50000 to 55000, inclusive.
NOTE: The Virtual ID must be unique for each patient monitor connected to the S/5 Network.
2.
Below
IX Network
, select
DHCP
or
Manual Configuration
.
a.
Enter a
Static IP
address.
b.
Enter a valid
Netmask
level.
c.
Enter a valid
Default Gateway
.
d.
Enter valid
DNS Server
addresses if applicable.
3.
Select
Speed & Duplex
option.
4.
Select
Save
.
The default network speed is 10 Mbps half duplex, which suppresses the effects of network
storms resulting from network misconfigurations. Please ensure that the network in your
facility meets the network configuration requirements as described in the Patient Monitoring
Network Configuration Guide.
Network connection(s)
Steps to be performed
S/5 Network only
1 and 4
IX Network only
2, 3, and 4
Both S/5 and IX Network
1, 2, 3, and 4
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...