Hardware installation
61
2106778-001
6.4.2 Installing a PSMP module to an F5 Frame
1.
Connect a PSMP module by aligning it with the insertion guides on the outside of the
frame. Push the module into the frame until it stops.
2.
To remove the PSMP module, pull the pull tab out and slide the module out of the guides.
6.4.3 Installing a PSMP module to an F7 Frame
WARNING
PHYSICAL INJURY- Take care when mounting devices to an IV pole. If a
device is mounted too high the IV pole may become unbalanced and tip
over.
1.
Connect the PSMP pole mount or frame mount cable to the back of the F7 Frame.
2.
Connect the PSM module to the mount.
6.4.4 Installing another module to a F5 or F7 Frame
WARNING
Ensure that the CARESCAPE respiratory modules are in vertical position
when used. Tilting them may result in erroneous readings.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......