CARESCAPE Monitor B850
84
2106778-001
7.9 Setting power frequency
WARNING
Incorrect power line frequency setting could adversely affect ECG and EEG
processing.
1.
Log in to Webmin.
2.
Select
Configuration
>
Power Frequency
.
3.
In the
Power Frequency
window, select the applicable power line frequency.
4.
Select
Save
.
The power frequency configuration takes effect immediately.
7.10 Selecting language and locale
Select the language used in clinical user interface and select the keyboard locale setting for the
alphanumeric keyboard and the barcode reader.
1.
Log in to Webmin.
2.
Select
Configuration
>
Language
.
3.
In the
Language
window, select the patient monitor language and keyboard language:
a.
Select the patient monitor language from the drop-down list and select
Save
.
b.
Select the keyboard locale from the drop-down list and select
Save
.
The language takes effect after the patient monitor is restarted. The keyboard locale takes
effect immediately after it is submitted.
7.11 Selecting national requirements
Activate France specific defaults for the ECG HR adjustment range and the reminder beep
behavior.
1.
Log in to Webmin.
2.
Select
Configuration
>
National Requirements
.
3.
In the
National Requirement
window, select the applicable option:
4.
Select
Save
.
The national requirements changes take effect immediately.
7.12 Configuring modules
You can configure some acquisition module specific settings. The settings are saved to the
permanent memory of the related acquisition module and the settings travel with the module
from one patient monitor to another.
These settings are pre-configured at factory, except the
Assets Settings
. You may need to
re-configure them after corrective maintenance, or for administration purposes.
Value
Description
None
Select the normal defaults.
France
Enable the following country specific monitoring:
- Heart Rate high alarm limit maximum 230.
- Reminder beep will sound every 2 minutes when alarms have
been silenced permanently.
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......