Configuration
91
2106778-001
activation process with a delay during an active patient case, but the new settings will activate
only after the next case reset / patient discharged.
1.
Log in to Webmin.
2.
Select
Configuration
>
Settings
.
3.
In the
Sub-Modules for Settings
menu, select
Activate
.
4.
In the
Activate Settings
window, select the settings that you want to activate:
•
Select
Cloned settings (clinical and platform)
to activate both the clinical and
platform settings.
•
Select
Cloned clinical settings
to activate only loaded clinical settings.
•
Select
Cloned platform settings
to activate only loaded platform settings.
NOTE: Activation of the cloned clinical or platform settings will not affect the following
settings in the target patient monitor:
Network
Unit and Bed Name
Host Asset Settings
Licenses
Remember to configure the unaffected settings manually in the target patient monitor.
•
Select
Factory default settings
to restore all platform and clinical settings back to
the factory defaults.
•
Select
US default clinical settings
to restore all clinical and platform settings back
to the US specific factory defaults.
NOTE: Activation of the factory defaults and US factory defaults will leave the following
settings in the target patient monitor unaffected:
- Host Asset Settings
- Licenses
5.
In the
Setting Activation
, select whether you want the new settings to take effect
immediately or after the next case end / discharge:
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......