CARESCAPE Monitor B850
142
2106778-001
Touchscreen
The quality of the display is not
good.
•
The cable is not connected
properly.
•
The video display is bad.
1. Check the connection between the
video display and the patient monitor.
If necessary, replace the video display
with a known good one.
2. If the problem is not corrected, replace
the processor board.
The power on/off switch on the
patient monitor does not
illuminate when it is toggled.
The power supply is bad.
Replace the power supply.
Continuous beeping alarm and
alarm light flashing yellow.
Mains supply is lost or the USB
cable is disconnected.
Restore the mains supply or reconnect the
USB cable.
Problem
Possible cause
Recommended actions
Touchscreen operation
inaccurate.
Touchscreen not calibrated.
Calibrate touchscreen, see section
.
Touchscreen inoperative.
Touchscreen USB cable loose
Connect the USB touchscreen cable to the
CPU unit or to the display.
NOTE: Secure the connected USB cable to
the CPU unit. Use the existing retaining
clips attached to the CPU unit.
Touchscreen USB cable
connected to the wrong USB
connector in the monitor.
Connect the touchscreen USB cable to the
correct connector.
Faulty touchscreen sensor.
Replace the display.
Problem
Possible cause
Recommended actions
Summary of Contents for CARESCAPE Monitor B850
Page 12: ...CARESCAPE Monitor B850 4 2106778 001 For your notes ...
Page 16: ...CARESCAPE Monitor B850 8 2106778 001 For your notes ...
Page 56: ...CARESCAPE Monitor B850 48 2106778 001 For your notes ...
Page 76: ...CARESCAPE Monitor B850 68 2106778 001 For your notes ...
Page 112: ...CARESCAPE Monitor B850 104 2106778 001 For your notes ...
Page 128: ...CARESCAPE Monitor B850 120 2106778 001 For your notes ...
Page 156: ...CARESCAPE Monitor B850 148 2106778 001 ...
Page 160: ...CARESCAPE Monitor B850 152 2106778 001 For your notes ...
Page 177: ...Service parts 169 2106778 001 13 3 Exploded view ...
Page 178: ...CARESCAPE Monitor B850 170 2106778 001 For your notes ...
Page 182: ...CARESCAPE Monitor B850 A 4 4 2106778 001 Notes Used service parts Signature Date ...
Page 186: ...CARESCAPE Monitor B850 B 4 4 2106778 001 For your notes ...
Page 190: ...CARESCAPE Monitor B850 C 4 4 2106778 001 For your notes ...
Page 191: ......