![GE AKTAprocess Operating Instructions Manual Download Page 69](http://html.mh-extra.com/html/ge/aktaprocess/aktaprocess_operating-instructions-manual_128019069.webp)
FH-231
HV-301
XV
-025
XV
-027
XV-072
XV-026
HV-303
HV-302
Open valve(s)
Valve positions
XV-026
Bypass (default)
XV-025 + XV-027
Inline
XV-025 + XV-027 + XV-072
Out_through_drain
HV-301 + HV-302 + HV-303 + XV-027 + XV-072
Drain (No flow)
The Drain valve position is used, for example, when the filter housing is emptied before
replacing the filter. The pump(s) must be set to 0.0 l/h when the Drain valve position is
used.
The UNICORN instruction ManFlow must be used when running
Out_through_drain unless the system has a flow meter before the air trap.
However, it is possible to use Flow for short periods of time even if the system
is not equipped with a flow meter.
Note:
ÄKTAprocess Operating Instructions 29-0252-49 AA
69
3 System description
3.4 Optional components
Summary of Contents for AKTAprocess
Page 1: ...GE Healthcare Life Sciences ÄKTAprocess Operating Instructions Original instructions ...
Page 2: ......
Page 179: ......