WARNING
Use columns that withstand expected pressures. If not, the columns
might rupture, resulting in injury.
WARNING
Never operate ÄKTAprocess with pressure control valves (PCVs)
completely closed. A pressure increase may result, and cause
leakage.
WARNING
Alarm signals. All alarm signals must be set within the limits
specified in the system documentation. Pressure and temperature
control must be activated while the system is in use to prevent the
piping system to leak or break.
WARNING
Alarms signals. Make sure to change back to the original alarm
level after UNICORN alarm buzzer test.
CAUTION
If an external Uninterruptible Power Supply (UPS) is used, this unit
must be powered before any other equipment.
CAUTION
To safely operate ÄKTAprocess, knowledge of how to use UNICORN
is required. Refer to UNICORN user documentation as required.
CAUTION
When handling the operator console, make sure that no body parts
are caught between the sections of the console arm.
ÄKTAprocess Operating Instructions 29-0252-49 AA
27
2 Safety instructions
2.1 Safety precautions