CAUTION
Do not insert your fingers or other objects into fans or other moving
parts.
CAUTION
Use ear protection whenever working close to the system in
operation.
CAUTION
Make sure that the pH electrode is mounted correctly after reassem-
bly.
Maintenance
WARNING
LOCK OUT / TAG OUT! Before any maintenance or decommission-
ing work is performed on ÄKTAprocess, make sure that:
•
it is empty and depressurized.
•
it is disconnected from process feed, electrical power and
pneumatic supply.
•
it is prevented from accidentally becoming re-energized during
maintenance.
•
it is clearly tagged as taken out of operation.
•
all process wetted areas are clean and decontaminated.
WARNING
Only personnel authorized by GE Healthcare may perform service,
installation, and maintenance of components inside the
ÄKTAprocess cabinet.
28
ÄKTAprocess Operating Instructions 29-0252-49 AA
2 Safety instructions
2.1 Safety precautions