B40/B20 Patient Monitor
15-4
Patient connections
1.
Connect the pressure transducer to the transducer adapter cable. Connect the cable to
the red connector in the module, or to the dual invasive blood pressure adapter cable.
NOTE: Invasive pressures need to be zeroed after reconnecting the pressure
transducer or cable, and whenever the patient’s position is changed. The transducer is
always leveled to the mid right atrium.
2.
Prepare the transducer kit according to the manufacturer’s instructions. Mount the kit
with the transducer zeroing port at mid-heart level.
3.
Ensure that there is no air in the line. Refer to transducer manufacturer’s instructions on
how to remove trapped air from the transducer.
4.
Connect the patient catheter to the pressure line.
5.
Open the dome stopcock to room air.
6.
Zero the transducer. See
"Starting with accurate values" on page 15-5
7.
Open the dome stopcock to pressure catheter and check the quality of the waveform.
Figure 15-3 Invasive blood pressure setup
(1) Compatible
InvBP
measurement capability
(2) Heparinized fluid bag with pressure infusor
(3) Flushing
set
(4) Disposable
catheter
(5) Transducer
(6) Adapter cable for the InvBP transducer
(7) Adapter cable for dual InvBP measurement
NOTE: Patient connections made according to the picture above using approved
accessories are defibrillator-proof.
3
5
4
7
2
6
1
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......