B40/B20 Patient Monitor
6-16
Configuring trends and snapshot
Changing default trend
You can select graphical or numerical trends to be displayed by default:
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Trends & Snapshot - Default Trend
Configuring trend pages
You can change the parameters on the trend fields:
1.
Press the
Monitor Setup
key.
2.
Select
Install/Service
and enter the password.
3.
Select
Trends & Snapshot - Graphical Trends
.
4.
Select the trend page that you want to change.
5.
Select graphical parameters for each field.
The field numbers start from the top of the screen. Select one parameter for each field on the
trend page, or turn the field
OFF
. When all the fields are
OFF
, the page is displayed with empty
fields. The time scale and page number appear at the bottom of the page.
Field 1-5 can be displayed on screen and Field 6 can be printed.
If several similar fields are selected on top of each other, they form one higher field. Equal fields
cannot be defined separate to each other.
NOTE: You cannot make changes in numerical trend page configuration.
Trends & Snapshot
Default Trend
Previous Menu
Graph
Change default trend type:
graphical or numerical.
Install/Service
Graphical Trends
Snapshot
Monitor
Setup
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......