B40/B20 Patient Monitor
6-14
Changing the recorder and printer settings
Recorder settings
1.
Press
Print/Record
.
2.
Select
Record Waveforms.
•
Select
Waveform 1
and select a parameter or select
OFF
. Then select
Waveform 2
and
Waveform 3
and their parameters. You can record up to three waveforms
simultaneously. If you wish to record only one waveform, select the other waveform
fields
OFF
.
•
Select
Start on Alarms
-
Yes
to start automatic strip chart recording when the
following alarms reach the red alarm level (
NO
is the default):
−
with VSP software license: bradycardia, tachycardia, asystole, Art high/low, V
Fib, V Tach
•
Select
Delay
and
OFF
or
12 s
. If the recording delay is
OFF
, the recording starts when
an event occurs and continues for 30 seconds or until manually stopped, or until the
recorder runs out of paper. If the delay is
12 seconds
(default), the recording starts
when an event occurs and the 12 seconds prior to the event are recorded from the
recorder memory. The recording continues for 18 seconds if the length has been set
to 30 seconds or until the recorder runs out of paper.
•
Select
Paper Speed
and
1
,
6.25
,
12.5
, or
25 mm/s
(default).
•
Select
Length
and choose
30 s
(default) or
Cont
.
Print/
Record
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......