Cleaning and care
9-1
9 Cleaning and care
Safety precautions
Warnings
•
Disconnect equipment from power line before cleaning
•
If liquid has accidentally entered the system or its parts, disconnect the power cord from
the power supply and have the equipment serviced by authorized service personnel.
•
Regular preventive maintenance should be carried out annually.
•
Do not use unspecified cleaners, materials or chemicals as they may damage device
surfaces, labels, or cause equipment failures.
•
Do not pour or spray any liquid directly on cables or leadwires or permit fluid to seep into
connections or openings.
•
Never use conductive solutions, solutions that contain wax, or wax compounds to clean
devices, cables or leadwires.
•
Do not immerse any part of the device in liquids or allow liquid to enter the interior.
•
The decision to sterilize must be made per your institution’s requirements with an
awareness of the effect on the integrity of the cable or leadwire.
•
Clean the surface of the probe before and after each patient use.
•
CABLE/SENSOR AFTER CARE
−
Do not immerse sensors or patient cables in water, solvents or cleaning solutions.
−
Do not reuse sensors intended for single patient use.
−
Do not sterilize sensors or patient cables by irradiation, steam, or ethylene oxide.
−
Clean the surface of the probe before and after each patient use.
−
Allow sensor and cable to dry completely after cleaning. Moisture and dirt on the
connector can affect the measurement accuracy.
−
If a probe is damaged in any way, discontinue use immediately.
−
Inaccurate SpO
2
data can result if a sensor is past its useful life.
−
A damaged sensor or a sensor soaked in liquid may cause burns during
electrosurgery.
•
Since calibration gas contains anesthetic agents, always ensure sufficient ventilation of
the room during calibration.
•
Inaccurate readings due to:
−
use of unapproved accessories
−
reuse of single-use accessories
−
affect on pressure strong scavenging suction
•
The user may only perform maintenance procedures specifically described in this
manual.
•
On the gas module indicates that airway gases should be calibrated every six months in
normal use and every two months in continuous use.
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......