B40/B20 Patient Monitor
2-18
Parameters specifications
ECG specifications
Leads available
3-lead configuration: I, II, III
5-lead configuration: I, II, III, aVR, aVL, aVF and VA
QRS detection range
0.5 to 5mV
QRS detection width (Q to
S)
40 to 120 ms
Defibrillation protection
5000 V, 360 J
Recovery time
<5 s
Input impedance
Common mode > 10 M
Ω
@ 50/60 Hz
Differential > 2.5 M
Ω
from 0.67 to 40 Hz
Common mode rejection
90 dB minimum at 50 Hz
Tall T wave rejection
>1.4 mV
ECG leads off detection
Active patient electrode: <30 nA
Reference electrode: <300 nA
Filter modes
50/60 Hz
Monitoring filter
0.5 to 40 Hz
ST filter
0.05 to 40 Hz
Dagnostic filter
0.05 to 150 Hz
Heart rate
Measurement range
30 to 300 bpm
Measurement accuracy
±5 % or ±5 bpm, whichever is greater
resolution
1 bpm
Heart rate response time (IEC 60601-2-27 201.7.9.2.9.101 b) 5))
Step increase from 80 to
120 bpm
average 6.9 s (6.5 to 7.5 s)
Step decrease from 80 to
40 bpm
average 8.2 s, (7.6 to 10.0 s)
The heart rate calculation operates with irregular rhythms of IEC 60601-2-27
201.7.9.2.9.101 b) 4), the heart rate after a 20 second stabilization period is:
Figure 3a
80 bpm
Figure 3b
59 bpm
Figure 3c
122 bpm
Figure 3d
117 bpm
Heart rate averaging computation (IEC 60601-2-27 201.7.9.2.9.101 b) 3)): Average
of 10 second median values
The average time and time range ( ) to alarm (VFib or VTachy) for tachycardia
waveform are as follows (IEC 60601-2-27 201.7.9.2.9.101 b) 6))
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......