Impedance respiration
12-3
Respiration rate source
Respiration rate is calculated automatically when ECG or CO
2
is measured unless the
respiration measurement is turned off. When CO
2
is measured, the respiration rate is
automatically calculated from CO
2
.
Respiration rate calculation switches back to impedance respiration if you press the
Silence
Alarms
key during an Apnea alarm.
The respiration rate source is displayed above the respiration numeric value.
Figure 12-2 Respiration rate source indication for impedance and CO
2
The impedance respiration waveform is displayed next to the RR value also when the RR value
is calculated from CO
2
.
To manually select impedance respiration as respiration rate source:
1.
Select respiration in a waveform or a digit field.
2.
Press the
Others
key.
3.
Select
Resp Setup
.
4.
Select
Resp Rate Source
and select
AUTO, CO2
or
Imped
.
Displaying impedance respiration
Figure 12-3 Impedance respiration waveform and numeric respiration rate value.
Others
Others
Resp Setup
Normal Screen
Set size , Resp rate source and
alarm. Enable or disable Resp
measurement.
Temp Setup
Entropy
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......