B40/B20 Patient Monitor
3-6
Installation checkout
It is recommended to be performed after monitor installation.
Skip the tests that are not
applicable for the installed monitor.
These instructions include a “
B. Installation and checkout form, B40/B20
” to be filled in when
performing the procedures.
Recommended tools
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for the
specified calibrations and adjustments to ensure accuracy.
NOTE: A functional tester cannot be used to assess the accuracy of pulse oximeter for monitor.
Table 3-1
Recommended accessories and tools
NOTE: For details on recommended accessories see “Supplies and Accessories“ catalog.
Visual inspection
Perform the following visual inspection to the installed monitoring system:
•
Carefully inspect the patient monitor if any damage.
•
Verify that the patient monitor is properly mounted with specified mounting solutions.
•
Verify that the cables between the patient monitor and the connected peripheral devices
are intact and properly connected to the right connectors.
Accessories
A rigid cylinder or pipe
NIBP cuff
Adult NIBP cuff hose with cuff ID
Infant NIBP cuff hose with cuff ID
Tubing parts to connect a manometer and a pump to
the NIPB cuff and hose.
Dual invasive pressure adapter cable
ECG accessories, IEC or AHA
- Multi-link 3-lead integrated cable and leadwire
- Multi-link 5-leadwire set
- Multi-link 3/5-lead ECG trunk cable
SpO
2
finger probe
SpO
2
Interconnect Cable
Temperature dual cable
Gas sampling line 3m/10 ft
Entropy cable
Tool
A multiparameter patient simulator with IBP, Temp
adpter cables
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......