![GE HEALTHCARE B20 User'S Reference Manual Download Page 194](http://html.mh-extra.com/html/ge-healthcare/b20/b20_users-reference-manual_571888194.webp)
B40/B20 Patient Monitor
11-12
Pacemaker detection
Safety precautions
Warnings
•
Do not diagnostically interpret pacemaker spike size and shape.
•
A pacemaker pulse can be counted as a QRS during asystole in either pace mode. Keep
pacemaker patients under close observation.
•
The shape of QRS complex may be changed so much because of the pacemaker that QRS
detection may be affected.
•
The monitoring of pacemaker patients can only occur with the pace program activated.
•
PACEMAKER PATIENTS - Rate meters may continue to count the pacemaker rate during
occurrences of cardiac arrest or some arrhythmias. Do not rely entirely upon rate meter
alarms. Keep pacemaker patients under close surveillance. See
for
the pacemaker pulse rejection capability of this instrument.
•
FALSE CALLS — False low heart rate indicators or false Asystole calls may result with
certain pacemakers because of pacemaker artifact such as electrical overshoot of the
pacemaker overlapping the true QRS complexes.
NOTES:
•
Pacemaker detection is always on.
•
The shape of QRS complex may be changed because of the pacemaker so much that QRS
detection may be affected.
•
If the patient has an atrial pacer, ST calculations can be performed if the pacer does not
coincide with the ISO point’s adjustment range.
•
Pacemaker detector may not operate correctly during the use of high-frequency (HF)
surgical equipment. The disturbances of HF surgical equipment typically cause false
positive pacer detection.
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......