System description
2-1
2 System description
Safety precautions
Warnings
•
All system devices must be connected to the same power supply circuit
•
EXCESSIVE LEAKAGE CURRENT - Do not use a multiple socket outlet or extension cord.
•
INTERFACING OTHER EQUIPMENT - Devices may only be interconnected with each other
or to parts of the system when it has been determined by qualified biomedical personnel
that there is no danger to the patient, the operator, or the environment as a result. In
those instances where there is any element of doubt concerning the safety of the
connected devices, the user must contact the manufacturers concerned (or other
informed experts) for proper use. In all cases, safe and proper operation should be verified
with the applicable manufacturer’s instructions for use, and system standards IEC 60601-
1 must be complied with.
•
Verify compatibility of all system components prior to installation.
•
Do not use multiple modules with identical measurements in the same monitor.
•
The parameter modules are not able to withstand unpacked drops from a height of 1 m
without damaging the module latches. If the device is dropped, please service the device
before taking it back into use.
•
The medical electrical equipment or medical electrical system may be interfered with by
other equipment, even if that other equipment complies with CISPR EMISSION
requirements.
•
EXPLOSION OR FIRE-Using non-recommended batteries could result in injury/burns to the
patients or users. Only use batteries recommended or manufactured by GE.
•
EXPLOSION HAZARD - not suitable for use in the presence of a flammable anaesthetic
mixture with air or with oxygen or nitrous oxide.
•
EXPLOSION HAZARD - Do not incinerate the battery or store at high temperatures.
•
Do not disassemble, open, or shred the battery; do not short circuit the battery pack;
charge only with internal charger.
Cautions
•
Do not dismantle the battery.
Principles of functions
The monitor is a semi-modular multi parameter patient monitor. The monitor is designed for
nurses and other healthcare providers who care for patients in critical care, perioperative care,
emergency care, general care and intra-hospital transportation environments.
The modular design makes the system flexible and easy to upgrade.
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......