![GE HEALTHCARE B20 User'S Reference Manual Download Page 197](http://html.mh-extra.com/html/ge-healthcare/b20/b20_users-reference-manual_571888197.webp)
ECG
11-15
for several minutes after the learning phase to allow the algorithm to reach optimal
detection performance.
About the arrhythmia detection
When an ECG signal is detected at the start of monitoring, the arrhythmia detection algorithm
begins acquiring and analyzing QRS complexes in the leads used for arrhythmia detection. This
phase is known as learning. Once learning is complete, the dominant QRS complex is stored as
a reference template. Reference template is used as a normal morphology of that patient and
it is compared with incoming beats to identify possible arrhythmias.
The EK-Pro arrhythmia detection algorithm is used. EK-Pro simultaneously analyzes leads I, II,
III, and V. Once learning is complete, the dominant QRS complex becomes the template.
The algorithm uses continuous correlation, incremental template updating and contextual
analysis. Continuous correlation attempts to find the best match between each incoming
complex and the set of stored (learned) templates. If no match is found with the existing
template, a new template is stored for the identified new QRS shape. Incremental template
updating allows information from each beat, that correlates over time, to be reflected in the
associated template. Contextual analysis uses information from neighboring QRS complexes
along with existing template measurements to make the best possible decision regarding the
beat’s origin (e.g., early, wide).
NOTE: With the 3 leadwire trunk cable, the algorithm uses the only one available lead
ECG1 Lead
, which is I, II or III, depending on the selected user lead.
Starting relearning manually
When the patient’s ECG pattern changes considerably, the monitor should start relearning a
new ECG pattern. The pattern changes, for example, when changing the patient’s position. To
start relearning manually:
1.
Press the
ECG
key.
2.
Select
Relearn
.
ECG
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......