Troubleshooting
10-13
Advanced NIBP troubleshooting
Symptom
Possible cause and solution
NIBP measurement does
not work or values seem
unstable
•
Check that cuff tubings are not bent, stretched, compressed or loose.
•
Prevent motion artifacts.
•
Use cuffs of correct size.
Why does the mean
value display while the
associated systolic and
diastolic values display
as - - -?
The following conditions may cause the mean value to display in the NIBP
parameter window while the associated systolic and diastolic values display as
- - -.
•
Very low systolic and diastolic amplitude fluctuations (e.g., patient in shock).
•
Very small difference between the mean and systolic pressure or the mean
and diastolic pressure.
•
Loss of system integrity (e.g., loose connections or worn parts).
Assess the patient and perform a visual inspection of the equipment to ensure
system integrity.
Why is the monitor re-
inflating the cuff
automatically?
The cuff target pressure must be higher than the patient’s systolic pressure to
obtain an accurate systolic and diastolic measurement. If a systolic blood
pressure cannot be found, the monitor searches for a systolic reading by re-
inflating the cuff at a higher pressure. This systolic search may occur once per
NIBP determination cycle. During a systolic search, the maximum cuff inflation
pressure will not exceed the normal pressure range of the cuff.
The monitor may be taking a control measurement. If the measured NIBP value
exceeds the alarm limits, a single low priority alarm sounds and a new
measurement is automatically taken. If the new value (the control
measurement) also exceeds the alarm limits the alarm priority escalates to
medium. In Manual mode and STAT mode there are at least four seconds
between the first measurement and the Control measurement for Adult and
Child cuffs, eight seconds for Infant cuffs. In Auto mode there are at least 30
seconds between the first measurement and the control measurement.
If
Weak pulsation
is detected, the module repeats the measurement up to
three times. Assess the patient and perform a visual inspection of the
equipment to ensure system integrity.
If cuff occlusion is detected during inflation the module might repeat the
measurement. Perform a visual inspection of the equipment to ensure system
integrity.
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......