B40/B20 Patient Monitor
2-20
GE TruSignal SpO
2
specifications
Measurement and display
range
1 to 100%
Calibrated against functional oxygen saturation.
Measurement accuracy
Adult/Pediatric
(100 to 70%):
without motion: ±2 digits (±3 digits with ear sensor)
with motion: ±3 digits
low perfusion: ±3 digits
Neonatal
(100 to 70%):
without motion: ±3 digits
with motion: ±3 digits
Adult/Pediatric/
Neonatal (69 to 1%):
unspecified
Display resolution
1 digit (1% of SpO
2
)
Display averaging
Normal 12s, Fast 3s
Wavelength of SpO
2
probe
LEDs:
Infrared LED 940 nm
RedLED 660 nm
Maximum energy of SpO
2
probe LEDs:
Infrared LED 42 µJ/pulse
Red LED 62 µJ/pulse
NOTE:
This information may be useful to clinicians, such as those performing photodynamic therapy.
Pulse rate
Measurement and display
range
30 to 250 bpm
Display resolution
1 bpm
Measurement accuracy
Without motion:
±2 bpm (Adult/Pediatric/Neonatal)
With motion:
±3 bpm (Adult/Pediatric/Neonatal)
Low Perfusion:
±5 bpm (Adult/Pediatric)
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......