B40/B20 Patient Monitor
10-8
Temp sensor error
Frame temperature sensor error.
•
Contact authorized service
personnel.
T1 temperature error
T2 temperature error
Hardware failure or unsuccessful
calibration.
•
Check the probe whether is set to
400.
•
Contact authorized service
personnel.
V Fib
Physiological alarm.
•
Check the patient status.
V Tach
Physiological alarm.
•
Check the patient status.
Weak pulsation
Weak or unstable oscillation signal
due to improper cuff position or
attachment, weak or abnormal blood
circulation, slow heart rate associated
with artifacts, moving or disturbed
patient during measurement, small air
leak.
•
Check the patient status.
•
Check the NIBP cuff position and
attachments.
•
Check that the cuff is not damaged.
•
Repeat the measurement and
check the patient status.
Zero ICP separately
The ICP channel must be zeroed
separately from all other invasive
pressures.
•
In IBP menu to zero ICP labelled
channel separately.
xxx high/low
Physiological alarm.
•
Check the patient status.
•
Adjust the alarm limits.
xxx measurement removed
Acquisition module has been removed.
•
Check and connect the module if
you want to restart the
measurement.
•
Check the patient status.
Message
Explanation
What to do
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......