![GE HEALTHCARE B20 User'S Reference Manual Download Page 288](http://html.mh-extra.com/html/ge-healthcare/b20/b20_users-reference-manual_571888288.webp)
B40/B20 Patient Monitor
17-20
Dips in capnogram
The dips seen in the capnogram during expiration are related to the sidestream gas sampling,
the continuous gas flow to the Y-piece, and patient’s cardiac contractions, which cause intra-
thoracic pressure changes and therefore flow variations.
The alterations in expired CO
2
waveform are cardiogenic movements of exhaled and circuit
gas at the sidestream gas sampling site. When the respiratory gas flow drops below the gas
sampling rate, a variable mixture of CO
2
free fresh gas and exhaled CO
2
rich gas is sampled.
This causes variations in sampled CO
2
concentrations.
In the illustration below, CO
2
waveform is the one on top, and flow is the lower waveform.
1.
Expiration
2.
Cardiogenic oscillations
Cardiogenic oscillations appear when:
•
A continuous fresh gas flow is fed into the patient Y-piece.
•
Sidestream gas sampling is done at the Y-piece.
•
The patient is ventilated with a long expiration time or low respiration times, and when
there is a long zero flow at end-expiration for some other reason.
Oscillations can be eliminated by adding a spacer with a 5 ml dead space between the Y-piece
and the airway adapter. Increased dead space creates a buffer volume between the Y-piece
and the sampling point, preventing the inspiratory and expiratory air from mixing during gas
sampling. Misinterpretation of EtCO
2
information can be avoided through identifying
cardiogenic oscillation and understanding the reasons for it.
EtCO2 value %
EtCO2 value mmHg
Indicates
4.5 to 5.5%
34 to 41 mmHg
normocapnia
< 4%
< 30 mmHg
hypocapnia
> 6%
> 45 mmHg
hypercapnia
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......