42
HiToP
®
184
gbo Medizintechnik AG
Version 2.0
8. If the data cannot be transmitted, the following note will be shown:
Menü
FIX
I
II
III
IV
Patient unit
Program
Repeat
Cancel
Please place the patient unit on the
side at a distance of 30 - 50 cm,
change it to the programming modus
and press the Softkey.
Figure 27:
Menu Programming of patient unit / Transmission error!
9. Correct the location of the patient device and press the
Softkey
[Repeat].
10. Upon correct data transmission the following note will be shown on the display:
Successful data transmission
(see Figure 26: Menu Programming of patient unit / Successful data transmission)
.
5.2.2 Selection of channel I, channel II, channel III , channel IV
Here you select the type of current and its parameters set in the respective channel. These can be
SimulFAM
®
i,
SimulFAM
®
X
or the
nerve block
. These parameters can be changed in a changing mask
before the programming.
1. Move with the
Modification knob
to the desired channel.
2. The type font will be shown inversely.
3. Press the
Softkey
[Select].
4. You move to the display of the preset type of current of the respective channel.
5.2.3 Selection SimulFAM
®
X
1. Move with the Modification knob to
SimulFAM
®
X
.
2. The selection is shown inversely.
3. Press the
Softkey
[Select].