18
HiToP
®
184
gbo Medizintechnik AG
Version 2.0
2.4.4 Modification of the basic settings
Indication Menu
selected in Unit settings:
1. Press the black
Menu key
.
2. You are in the display window Main menu.
3. Move the scroll bar with the
Modification knob
to Basic settings.
4. Confirm by pressing the white
Softkey
[Continue].
You are in the menu Basic settings. The menu provides the following options:
Basic settings
Setting possibilities
Delivery
Scan – Starting frequency
= lower cut-off frequency for
SimulFAM
®
i
4096 - 16384 Hz
4096 Hz
Scan – Ending frequency
= upper cut-off frequency for
SimulFAM
®
i
8192 - 32768 Hz
32768 Hz
Treatment time FAM X
= basic setting of the therapy time which is preset upon
switching-on the device or selecting a channel.
5 – 90 min
60 min
Treatment time FAM i
= basic setting of the therapy time which is preset upon
switching-on the device or selecting a channel.
5 – 90 min
60 min
Slope FAM X
= basic setting of slope for
SimulFAM
®
X
which is preset
upon switching-on the device or selecting a channel.
10 – 200 %
150 %
Blocking frequency
= basic setting of the frequency for nerve block which is preset
upon selecting the blocking from the Indication menu.
4096 – 12274 Hz
12274 Hz
Intensity monitoring
=
see chapter 2.5 Intensity monitoring
on / off
on
Table 4:
Basic settings (Indication Menu)
5. Move the scroll bar with the
Modification knob
to the basic setting to be changed.
6. Confirm by pressing the white
Softkey
[Continue].
7. Move the cursor or change the value with the
Modification knob
to the desired value.
8. Confirm by pressing the white
Softkey
[Store].
9. Exit the device setting menu by pressing the black
Menu key
[Return].
10. On the display the Main menu appears again.
11. Exit the Main menu by pressing the black
Menu key
[Return].
12. On the display the Overview window appears again.