44
HiToP
®
184
gbo Medizintechnik AG
Version 2.0
Selection
Presetting
Setting possibilities
treatment time
from basic setting
1 min. - max. 90 min.
slope
from basic setting
10 - 200 %
pulse rise time
0 s
0 - 60 s
pulse holding time
0 s
0 - 60 s
pulse pause
0 s
0 - 60 s
number of treatments
20
1 – 50, unlimited
Table 8:
Setting possibilities
SimulFAM
X
5.2.4 Selection nerve block
1. Move with the
Modification knob
to nerve block.
2. Press the white
Softkey
[Select].
3. You move to the display nerve block:
Menü
FIX
Nerve block
Period of treatment
Frequency
16384 Hz
30 Jours
I
II
III
IV
Change
Patient unit
Program
Confirm
Return
Figure 30:
Menu Programming of patient unit / nerve block
Selection
Presetting
Setting possibilities
frequency
from basic setting
4096 – 12274 Hz
period of treatment
30 days
1 - 30 days
Table 9:
Setting possibilities nerve block
4. Move the scroll bar with the
Modification knob
to the setting to be changed. This will be shown
inversely.
5. Press the white
Softkey
[Modify].
6. The presetting will be flagged inversely.
7. Set with the
Modification knob
the desired value.
8. Confirm by pressing the
Softkey
[Modify] and you will remain on the same display to realize further
settings, or
9. confirm by pressing the white
Softkey
[Confirm]
and you exit the d
isplay Therapy II to move to display
Therapy III or exit display Therapy III to move to the transmission start window (
see Figure 25: Menu
Programming of patient unit / start).