2 Start of operation
2.1 Transport and installation
The High Tone Power Therapy device is a portable unit. In the base plate there is a carrying handle in the
form of a recessed grip. Each plane surface is appropriate to place the unit,. A wall distance of at least 20 cm
has to be observed. The device must not be placed in front of radiators.
The device for High Tone Therapy corresponds to the regulations EN 60601. It is a device of protection class
I. Within the scope of the MDD the High Tone Power Therapy device belongs to class IIa
(see also chapter
For cleaning and disinfection of
the current stimulation device and its accessories do not use any agents
containing higher portions of phenol derivatives, alcohol, compounds of chlorine or peracetic acid. It is
recommended to use disinfectants on aldehyde basis. Never use an abrasive. Fingerprints on the display
should be cleaned with a dry cloth or some isopropanol.
Warning!
Unplug the mains plug out of the socket prior to cleaning and disinfection.
Do not clean the device under running water and do not use liquid abrasive cleaner!
The device is not suited for heat sterilization or for the sterilization with gases.
The
HiToP
184
is suited for wiping disinfection. Make sure that no liquids soak into the device. Under no
circumstances the plug or socket must get wet. Do not sprinkle the device for cleaning or disinfection. The
device is not suited for hot sterilisation or sterilisation with gases.
2.1.1 Cleaning the Elastic Straps
The elastic straps can be washed in a washing machine. Please observe the washing instructions of the
instruction label.
2.1.2 Cleaning the Electrodes
You can clean the rubber electrodes after a treatment with some warm water (appr. 40°C) and a soap suds on
the conductive (black) side. Dry the electrodes with a cleaning cloth or let them air-dry. The cleaning process
removes the film left by the Aloe spray or the contact spray and the electrodes will regain a good
conductivity.
!!
Note!
The conductivity of the electrodes is optimized by the use of a certain amount of graphite.
Consequently, when using and cleaning the electrodes, black color may come off.
2.1.3 Disinfecting the Electrodes