HiToP
®
184
gbo Medizintechnik AG
Version 2.0
43
4. The
SimulFAM
®
X
window opens on the display:
Menü
FIX
3 Hz
10 Hz
20 Hz
100 Hz
1,5 - 10 Hz
10 - 20 Hz
20 - 200 Hz
SimulFAM X
0,1 - 1,5 Hz
I
II
III
IV
Patient unit
Program
Select
Return
Figure 28:
Menu Programming of patient unit / SimulFAM
X
5. Select the desired
SimulFAM
X
frequency with the
Modification knob
.
6. Press the
Softkey
[Select].
7. You move to the display
SimulFAM
X
, example here: 1.5 - 10 Hz.
Menü
FIX
SimulFAM X 1,5 - 10
Treatment time
60 min
Slope
74 %
Pulse - Rise time
2 s
Pulse - Holding time
20 s
Pulse - Pause
Number of treatments
20 s
10
I
II
III
IV
Change
Patient unit
Program
Confirm
Return
Figure 29:
Menu Programming of patient unit / modify the settings
8. Move the scroll bar with the
Modification knob
to the setting to be changed. This will be shown
inversely.
9. Press the white
Softkey
[Modify].
10. The presetting will be flagged inversely.
11. Set the desired value with the
Modification knob
.
12. Confirm by pressing the
Softkey
[Modify] and you will remain on the same display to realize further
settings, or
13. Confirm by pressing the white
Softkey
[Confirm
] and you will exit d
isplay Therapy II and move to
display Therapy III or exit display Therapy III and move to the transmission window (
see Figure 25:
Menu Programming of patient unit / start
).