3.2.9 Keep therapy frequency constant
Figure 11:
FIX key
In
SimulFAM
X
it is possible with that key to stop the modulation frequency travers during the treatment
with frequency-modulated types of current and to release it by pressing the key again. After having stopped
the modulation frequency travers it is possible to select a fixed stimulation frequency with the
Modification
knob
.
In
SimulFAM
i
it is possible with that key to stop the frequency travers during the treatment and to
release it by pressing the key again. After having stopped the frequency travers it is possible to select a fixed
frequency on the threshold curve with the
Modification knob
.
3.2.10 Therapy time
05:00
min. sec.
Figure 12:
Therapy time key
Depending on the type of current the therapy time defined in the basic settings will be suggested. To change
the therapy time proceed as follows:
1. Press
Function key Time
.
2. The therapy time will be shown inversely.
3. Set the desired period of time with the
Modification knob
.
4. The time can be modified by steps per minute. The maximum therapy time to select is 90 minutes.
Additionally, it is possible to select endles
s
; in this case there is no control of the therapy time through
the device.
!!
Note!
The selected or given therapy time passes only if the
Intensity regulator
is turned up from its
„0.0“
position.
The therapy time can be modified during the treatment.
At the end of the therapy time the stimulation current to the patient will automatically be
reduced:
The triad gong will sound
(see also Table 2: Acoustical user support).
The
Intensity regulator
automatically reverts to
„0.0“
-after switching off the stimulation
current to the patient.
FIX