39
5 Programming of patient unit
It is possible to load a maximum of three types of therapy current into the patient device. These can be
SimulFAM
®
i
, a certain
SimulFAM
®
X
frequency or the
nerve block
. The therapies are called Therapy I,
Therapy II and Therapy III in the patient device. In the following it is described how to program it. Place the
patient device with a distance of about 20 – 30 cm to the right of the device
HiToP
184
.
Change the patient device to the programming modus as follows:
1. Hold both
Function keys
+
and
–
pressed simultaneously.
2. Connect the device through the mains switch at the back of the device.
3. The display of
HiToP
181-H
shows the following:
Therapy survey
1. : I
00 Min.
2. : X 00 Min.
3. : B
0 days
4. The patient device may now be programmed.
5.1 Therapy I
Therapy I will always be allocated with the type of current
SimulFAM
®
i
. Only the therapy time and the
number of treatments can be changed before programming the patient device.
1. Press the black
Menu key
.
2. You are in the display window Main menu
(see Figure 3: Main menu).
3. The scroll bar has been preset to Program patient unit.
4. Confirm by pressing the white
Softkey
[Continue].
5. You are in the Menu Program patient unit Therapy I
SimulFAM
®
i
.
Menü
FIX
SimulFAM i
Therapy I
Number of treatments
10
Patient unit
Program
Change
Confirm
Return
I
II
III
IV
Treatment time
60 min
Figure 23:
Programming of patient unit with
SimulFAM
®
i