32
HiToP
®
184
gbo Medizintechnik AG
Version 2.0
7. Confirm by pressing the white
Softkey
Continue
.
8. The notes about the electrode application corresponding to the indication are shown on the display.
9. Apply the electrodes as indicated if there are no contraindications.
10. Confirm by pressing the white
Softkey
Continue
.
11. The types of current selected through the indication are shown on the display:
-
in case of
SimulFAM
i
advance the
Intensity regulator
and set the intensity values for the lower
and upper cut-off frequency. The therapy starts.
-
in case of
SimulFAM
X
advance the
Intensity regulator
. The detail window will be automatically
displayed and the therapy starts.
-
In case of nerve block advance the
Intensity regulator
up to the tolerance limit (no muscle
contractions!). The detail window will be automatically displayed and the therapy starts.
-
In case of fixed frequency advance the
Intensity regulator
until a minimal threshold has been
achieved. The therapy starts.
12. Upon travers of the first type of current the triad gong sounds.
13. Attention: if that blinking signal appears on the display
the therapy has not finished yet.
!!
Note!
Here it is possible to escape the programmed therapy by selecting another type of current.
14. The next type of current is shown on the display.
15. Advance the
Intensity regulator
and the therapy continues.
16. Upon travers of the second current frequency the triad gong will sound.
17. Proceed as described in point 13 - 17 until the end of the therapy.
18. At the end of the therapy time the triad gong will sound.
!!
Note!
In the indication menu up to 4 current types are offered for one therapy.
In the overview window the actual current type is indicated as follows:
Menü
FIX
12:22
P 1/3
13:25
30:22
SimulFAM
i
100 Hz
8192 Hz
3 Hz
26 V
12,5 V
0,0 V
Übersicht
125,6 mWh
I
II
III
IV
Ind.
Menü
Detail
Anzeige
Detail
Anzeige
Figure 21:
Therapy program in the indication menu
Therapy time of the
actual current type
1st current type of 3
actual current type