46
HiToP
®
184
gbo Medizintechnik AG
Version 2.0
6.2 Electrode positioning
The positioning of the patient is very important. The therapy shall always be carried out in a comfortable and
relaxed position. The joints shall be located in an angular position, so that they are placed in a mid-position,
from which both the flexor as well as the extensor muscles can be stimulated.
In general, the preparation and application of the electrodes is carried out as follows:
1. Connect the electrodes to the patient cable.
2. Connect the patient cable to the distribution cable.
3. Plug in the patient plug of the distribution cable to the referring channel socket of the unit.
4. Moisten the conductive rubber electrodes with contact spray to avoid current sensations on the skin.
5. If you want to use the macroporous intermediate layers or fleece pockets, they must be thoroughly
washed under running water.
6. Apply the electrodes to the patient.
7. Apply the elastic tape with closure in a way that the electrode fits completely.
Warning!
Do not apply the electrodes on skin injuries. Even minor abrasive burns can cause a
burning sensations to the patient. Herewith the intensity of the current can be
erroneously judged. If this cannot be avoided, apply zinc ointment or vaseline to cover
the affected parts of the skin.
The electrodes must rest on the skin with its complete surface in order to avoid
excessive local current densities. Otherwise there can occur skin redness, and
sometimes even burns.
It is recommended not to exceed the current density of 2 mA/cm
2
for all electrodes.
During the therapy the patient must take off jewelry and glasses.