1.8 Short instructions SimulFAM
®
X
1. Connect the device through the mains switch at the back of the device. After that the device carries out
an automatic check of all functions. The faultless automatic check ends with the acoustic signal (triad
gong).
2. Connect the accessories with the according socket on the right side of the device.
3. The general window with the presetting
SimulFAM
®
i
for all plugged channels is shown on the display.
4. After the automatic check, the
Softkeys
[Ind.Menü] as well as the
Channel keys
are active (constantly
lit). The CD module is active.
5. The black
Softkey
Menu is active. It is used for calling the main menu. There you can modify the
presettings.
6. Apply the electrodes to the patient
(see chapter
Fehler! Verweisquelle konnte nicht gefunden werden.
Fehler! Verweisquelle konnte nicht gefunden werden.
)
.
7. Select a channel with the
Channel key
.
8. The activated channel is shown inversely.
9. Select a current type of
SimulFAM
®
X
by pressing the
keys
on the left of the display.
10. The basic setting of the current type is set automatically.
!!
Note!
If the intensity regulator is not advanced within the next 10 sec. and/or another type of current is
selected, the inverse display will be extinguished. After advancing the intensity regulator the
following will appear:
Select channel
Activate the desired channel again.
11. Increase the current slowly with the
Intensity regulator
until the desired stimulation effect occurs.
12. The detail window is displayed automatically and the therapy starts. The bar display
SimulFAM
®
X
shows the frequency travers optically.
13. In case of traversing frequencies the present stimulation frequency is shown. The frequency values are
increased and reduced every second.
14. Within 20 seconds after having activated the intensity regulator, the program returns to the overview
window. Here the frequency and the voltage for the chosen channels are shown.
15. At the end of the therapy the triad gong will sound. The current to the patient will be automatically
reduced.