40
HiToP
®
184
gbo Medizintechnik AG
Version 2.0
6. You can keep the presettings with the
Softkey
[Confirm] and move to display Therapy II
(see chapter
5.2 Therapy II / Therapy III)
, or
7. move the scroll bar with the
Modification knob
to the setting to be changed. It will be shown inversely.
8. Press the white
Softkey
[Change].
9. This presetting will be flagged inversely.
10. Set the desired value with the
Modification knob
.
11. Confirm by pressing the
Softkey
[Store] and you remain in the same display to continue with further
settings, or
12. confirm by pressing the white
Softkey
[Confirm]
and you move to the di
splay Therapy II
(see chapter
5.2 Therapy II / Therapy III).
Selection
Presetting
Setting possiblities
therapy time
from basic setting
1 min. – max. 90 min.
number of treatments
20
1 – 50, unlimited
Table 7:
Menu Programming of patient unit
!!
Note!
Due to software version 2.02 the functionality for programming the
HiToP 181-H
is made more
flexible:
- The number of treatments increased from 20 to 50.
- Additionally you can set the number of treatments to unlimited. This function is realized in
HiToP
181-H
software version 1.06.
The procedure of programming will not be changed.
5.2 Therapy II / Therapy III
Therapy II and Therapy III can be programmed either with the set types of current and parameters of
channels I – IV or with the types of current
SimulFAM
®
i,
SimulFAM
®
X
or
nerve block
.
1. Upon confirming with the white
Softkey
[Confirm] of Therapy I you move to the display window
Therapy II:
Menü
FIX
Therapy II
Channel I
without
Channel III
Channel IV
SimulFAM X
SimulFAM i
Nerve block
Channel II
I
II
III
IV
Patient unit
Program
Select
Return
Figure 24:
Menu Programming of patient unit / Therapy II