Precautions about this Device
W A R N I N G
Use only the accompanying 3-way AC power cable. Use of other
cables may result in electric shock to the patient and the operator.
The power cable must be connected to AC 100–240V outlet.
Do not use multi-tap.
When using multiple ME equipment simultaneously, perform
equipotential grounding to prevent potential difference between the
equipment. Even a small potential difference may result in electric
shock to the patient and the operator.
Use the trolley only with the equipment specified by Fukuda Denshi.
Otherwise, the monitor and trolley may fall down, resulting in injury or
damage to the monitor.
Be sure to lock both casters when using or storing the trolley.
The trolley may move or fall down, resulting in injury or damage to the
monitor.
Do not use or store the trolley where it will be subject to inclination of
10 degrees or more. The trolley or equipment may fall resulting in
injury or damage to the equipment.
When the air filter is washed with neutral detergent, dry it completely
before reattaching. If the moisture is remained on the air filter, it may
damage the equipment.
The air filter must be attached after cleaning/replacing. If the
equipment is used with the air filter detached, it may damage the
equipment.
C A U T I O N
The HR-500 Recorder Module can record only waveforms (Manual
Rec., Alarm Rec., Periodic Rec.). Graphic recording and tabular
recording cannot be performed.
The internal switch setup will be performed by our service
representative. Users should not open the maintenance cover.
Do not insert unspecified modules into the Input Box. Otherwise, the
equipment may be damaged, and safety cannot be ensured.
If not using for a long period of time, turn OFF the power of the main
unit, Super Module, and Input Box.
As the Super Module and DSC-7300 communicates via Input Box, the
power of the Input Box must be always turned ON even if the module is
not inserted in the Input Box.
Clean the equipment frequently so stains can be removed easily.
To prevent injury, it is recommended to wear gloves when cleaning the
equipment.
Do not allow cleaning solution enter the connector part.
Do not use organic solvents, thinner, toluene and benzene to avoid
damaging the resin case.
Do not polish with abrasive or chemical cleaner.
When sterilizing the entire room using a spray solution, pay close
attention not to have liquids get into the connectors.
Use only neutral detergent to clean the housing. Do not use chemical
cloth, scrub brush, abrasive, polishing powder, hot water, volatile
solvent and chemicals (cleanser, thinner, toluene, benzine, benzol, and
synthetic detergent for house and furniture), or sharp-edged tools. The
surface resin coating may be damaged, resulting in discoloration,
scratches, and other problems.
Do not open the housing.
Do not allow alcohols or other liquids enter the equipment.
Use only the accessories specified for this device. Otherwise, proper
function cannot be executed.
For quality improvement, specifications are subject to change without
prior notice.
vi
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...