The input box data is not displayed on the monitor.
The “IB-7300 not configured” message is displayed on the monitor.
Cause 1: The input box setup is not properly performed.
Solution:
On the input box setup of the DS-7300 system, set the corresponded module for each
slot. For procedures, refer to the operation manual of the DS-7300 Bedside Monitor.
Cause 2: The DS-7300 and HS-700 are running an old program version.
Solution: Update the DS-7300 and HS-700 program version to V03-01 or later.
The “Check IB-7300 cooling fan” message is displayed on the monitor.
Cause
: The cooling fan is conjested reducing the cooling effect.
Solution
: Clean or replace the air filter.
The “IB-7300 hardware error” message is displayed.
Cause
The Input Box hardware is malfunctioning.
Solution
・
Immediately turn off the power and stop using the device.
・
Display the IB-7300 maintenance screen in the DS-7300 maintenance screen and
check the cause of the error.
・
Check the error status display on the back of the IB-7300.
・
Perform a memory test and check the cause of the error.
8
-
3
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...