No.
Check Item
Check Procedure
Criteria
14. Electrical Safety (
※
)
Perform the following check item as appropriate. Check these items when you have disassembled the equipment to
check / replace the boards or units.
08 Protective
Ground
Resistance
Measure the protective ground continuity
resistance.
According to test procedure of IEC 60601-1 18.
Should be 0.1
Ω
or below
between the protective
ground terminal and
accessible conductive part.
Protective ground resistance
of the power cable should be
0.1
Ω
or below.
09 Withstand
Voltage
Test (A-a1)
Apply AC1500V for 1 minute between
primary power source and accessible
metal part with protective ground (A-a1).
According to test procedure of IEC 60601-1 20.
Should withstand applied
voltage.
10 Withstand
Voltage
Test (A-a2)
Apply AC4000V for 1 minute between
primary power source and exterior
without protective ground (A-a2).
According to test procedure of IEC 60601-1 20.
Should withstand applied
voltage.
11 Withstand
Voltage
Test (B-a)
Apply AC4000V for 1 minute between
primary power source and applied part
(B-a).
Note: The internal protective circuit
composition differs according to
the equipment.
According to test procedure of IEC 60601-1 20.
Should withstand applied
voltage.
12 Withstand
Voltage
Test (B-d)
Apply AC1500V for 1 minute between
the exterior and applied part (B-d).
Note: The internal protective circuit
composition differs according to
the equipment.
According to test procedure of IEC 60601-1 20.
Should withstand applied
voltage.
10
-
19
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...