Daily Check List (HC-500)
Daily Check List
No.
Inspected
Date
Inspected
by
Location
Device Type HC-500
Serial No.
Date of Purchase
Item
Details
Criteria
Judgment
Appearance
Make sure to check visually that flaws,
cracks, deformation, and rust are not
present on the cabinet.
The defects shall not affect the
monitoring operation.
OK / NG
Make sure that the unit is installed in a
stable place.
The unit must not be installed on an
inclined surface, and must be
installed in a place free from
vibrations and impact.
OK / NG
Installation
Make sure that the unit is installed in a
place free from unwanted influences.
The environmental conditions of the
installation place must comply with
specifications including atmospheric
pressure, temperature, and humidity.
The unit must be installed in a place
where it will not be exposed to water
or any other liquid.
OK / NG
Zero
Calibration
Make sure that the zero calibration is
complete.
The Zero lamp is OFF.
OK / NG
Airway
Adapter
Calibration
Make sure that the airway adapter
calibration is complete.
The Airway lamp is OFF.
OK / NG
Function
Turn ON the patient monitor and make
sure that the unit is operating properly.
The CO
2
waveforms and
measurements must be displayed on
the monitor.
OK / NG
Periodical
Check
Check up the date on which the
periodical check was performed last
time.
It should be within 1 year.
OK / NG
DC fan
Check the date of the first use (the date
of exchange).
Used from ___ / ___ / 20___
Should be within 5 years after the
first use.
OK / NG
Comment
10
-
8
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...