Periodic Check List (HC-530)
Periodic Check List
No. .
Item
Inspection
Method / Criteria
Judgment
Comment
1. tcp Electrode
No damage should be found.
OK
/
NG
2. Accessories for Electrode
Check storage conditions.
OK
/
NG
Cords,
Accessorie
s
3. Operation Manual
Must be kept as specified.
OK
/
NG
1. Flaws, Cracks, Deformation and
Rust
Must be free from damage.
OK
/
NG
2. Panel and Label
Must not be damaged,
scratched or erased.
OK
/
NG
Ap
p
earan
c
e
Appearanc
e
3. Key, LED and Other
Must be free from damage.
OK
/
NG
1. Key
Must operate property.
OK
/
NG
Module
2. Input Connector
Must perform normally.
OK
/
NG
Mech
an
ica
l
Installation
into the
Input Box
1. Lock
Must perform normally.
OK
/
NG
1. CAL LED
Must light when the cable is
connected and electrode is
attached to the socket.
OK
/
NG
2. Calibration
The calibration must be
properly performed by pressing
the CAL key on the HC-530.
OK
/
NG
3. Electrode Attachment Time
After calibration, remove the
electrode from the socket and
set the electrode site time to 30
min. “SITE TIME OVER”
must be displayed when 30
minutes have elapsed.
OK
/
NG
4. Electrode Heating Control
The accuracy must be within
±
0.1
°
C.
OK
/
NG
Electrical
Function
5. Cable Detection
The data must disappear when
the cable is unplugged.
OK
/
NG
1. Enclosure Leakage Current (N.C.)
Less than 0.1mA
OK
/
NG
2. Enclosure Leakage Current
(S.F.C.)
Less than 0.5mA
OK
/
NG
3. Patient Leakage Current I (N.C.)
Less than 0.01mA
OK
/
NG
4. Patient Leakage Current I (S.F.C.)
Less than 0.05mA
OK
/
NG
Leakage
Current
5. Patient Leakage Current III (S.F.C.)
Less than 0.05mA
OK
/
NG
1. Between Isolated Applied Part and
Primary
AC 4kV/1min
OK
/
NG
Safety
Withstandi
ng Voltage 2. Between Isolated Applied Part and
Enclosure
AC 1.5kV/1min
OK
/
NG
Overall
Judgment
OK
/
NG
Equipment
tcpO
2
/pCO
2
Module
Model
Type
HC-530
Serial
Number
Installation
Place
Date of
Purchase
Date of
Inspection
Inspector
Approval
10
-
26
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...