HB-500 Invasive Blood Pressure Module
Size
90.0(W)
×
150.0(D)
×
30.0(H) mm (approximate)
Weight
300 g
Module slot
1 slot
(
when inserted into IB-7300 the input box
)
Environmental Conditions
Operating Temperature
:
10 ~ 40
°
C
Operating Humidity
:
30 ~ 85% (No Dew Condensation)
Operating Atmospheric Pressure
:
700
~
1060hPa
Storage Temperature
:
−
10 ~ 60
°
C
Storage Humidity
:
10 ~ 95% (at 60
°
C temperature, No Dew Condensation)
Storage Atmospheric Pressure
:
700
~
1060hPa
Safety
General Standard : IEC60601-1:1988 / EN60601-1:1990
(Medical electrical equipment – Part 1: General requirements for safety)
Amendment A1 to IEC 60601-1:1991 / EN60601-1:1993
Amendment A2 to IEC 60601-1:1995 / EN60601-1:1995
EMC Standard
: IEC60601-1-2:2001 / EN60601-1-2:2001
(Medical electrical equipment – Part 1: General requirements for safety –
2. Collateral standard: Electromagnetic compatibility – Requirements and
tests)
Classification
The class of protection
against electric shock
: Class I ( with the DS-7300 system)
The type of protection
against electric shock
: type CF
Waterproof Level
: IPX0 (no protection)
Disinfection Method
: Cleaning only
Degree of safety in the presence of flammable anesthetic mixture:
not suitable for use in the presence of flammable anesthetics mixture
with air or with oxygen or nitrous oxide.
Operation Mode
: Continuous Operating Equipment
Power Requirements
Voltage
:
24 VDC
±
10%
Power Consumption
:
6.0 VA (max.)
Measured Parameters
Invasive Blood Pressure (2 Channels)
Operation Keys and Indicators
BP_A
key
(
BP_A Zero Balance
)
, BP_A Zero Lamp
BP_B
key
(
BP_B Zero Balance
)
, BP_B Zero Lamp
Functions
Host Program Boot Function ( From Monitor to Flash Memory)
2
-
4
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...