Periodic Check List (HB-500)
Periodic Check List
No. .
Item
Inspection
Method /Criteria
Judgment
Comment
1.Interface Cable
Check visually, and by cable
conduction test
OK / NG
2.Channel Identification Label
Check storage conditions.
OK / NG
Cables,
Consumables,
Accessories
3. Operation Manual
Should be stored in specified
location.
OK / NG
1.Flaws, Cracks, Deformation and
Rust
Must be free from damage.
OK / NG
2.Panel and Label
Not be damaged, scratched or
erased
OK / NG
Ap
p
earan
c
e
External
Appearance
3.Key, LED and Other
Must be free from damage.
OK / NG
1.Key
Must operate property.
OK / NG
Mech
an
i
ca
l
Module
2.Input Connector
Must perform normally.
OK / NG
1. Zero Balance
Must is complete
with COBE simulator.
OK / NG
2. Sensitivity
Must be accuracy within
±
2mmHg At 100mmHg with
COBE simulator.
OK / NG
Electrical
Function
3. Cable Detection
Must disappear BP waveform
and value when unplugging
cable.
OK / NG
1.Enclosure Leakage Current
(N.C.)
Less than 0.1mA
OK / NG
2.Enclosure Leakage Current
(S.F.C.)
Less than 0.5mA
OK / NG
3.Patient Leakage Current
Ⅰ
(N.C)
Less than 0.01mA
OK / NG
4.Patient Leakage Current
Ⅰ
(S.F.C.)
Less than 0.05mA
OK / NG
Leakage
Current
5.Patient Leakage Current
Ⅲ
(S.F.C.)
Less than 0.05mA
OK / NG
Safety
Withstand
Voltage
1.Between Isolated Applied Part
and Enclosure
AC1.5kV/1min
OK / NG
Overall Judgment
OK / NG
Equipment
Invasive Blood Pressure
Module
Model
Type
HB-500
Serial
Number
Installation
Place
Date of
Purchase
Date of
Inspection
Inspector
Approval
10
-
23
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...