Compliance to the Electromagnetic Emissions
The DS-7300 system is intended for use in the electromagnetic environment specified below.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF Emissions
CISPR 11
/EN55011
Group 1
The equipment uses RF energy that is necessary for the
internal functioning of the equipment itself. Therefore, its
RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
/EN55011
Class A
This equipment is suitable for use in all establishments other
than domestic and those directly connected to a low-voltage
power supply network which supplies buildings used for
domestic purposes.
Compliance to the Electromagnetic Immunity (1)
The DS-7300 system is intended for use in the electromagnetic environment specified below.
It should be assured that the DS-7300 system is used in such an environment.
Immunity Test
IEC60601-1-2
Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC/EN61000-4-2
±
2, 4, 6kV
contact
±
2, 4, 8kV
air
±
2, 4, 6kV
contact
±
2, 4, 8kV
air
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Electrical fast
transient / burst
IEC/EN61000-4-4
±
2kV
for power
supply lines
±
1kV
for
input/output lines
±
2kV
for power
supply lines
±
1kV
input/output
lines
Power supply quality should be at
levels characteristic of a typical
location in a typical commercial or
hospital environment.
Surge
IEC/EN61000-4-5
±
0.5, 1kV
:
differential mode
±
0.5, 1, 2kV
:
common mode
±
0.5, 1kV
:
differential mode
±
0.5, 1, 2kV
:
common mode
Power supply quality should be at
levels characteristic of a typical
location in a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines.
IEC/EN61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(95% dip in U
T
)
for 5sec.
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(95% dip in U
T
)
for 5sec.
Power supply quality should be at
levels characteristic of a typical
location in a typical commercial or
hospital environment.
If it is required to continuously
operate the DS-7300 system during
power failure, it is recommended to
operate on an uninterrupted power
supply.
Power Frequency
(
50/60Hz
)
Magnetic Field
IEC/EN61000-4-8
3A/m 3A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Note :
U
T
is the AC mains voltage prior to application of the test level.
viii
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...