No.
Check Item
Check Procedure
Criteria
9. Invasive Blood Pressure
01
BP Accuracy
Input 0, 100, 200mmHg with
comprehensive tester, and measure
the error between the displayed value
and input value.
Error should be within display
full scale (300mmHg) ±2% or
±1mmHg whichever larger.
02 Pulse
Detection
Accuracy*
Input 30, 120, 160bpm with
comprehensive tester, and measure
pulse rate error.
Pulse rate error should be
within ±3%.
03 BP
Probe-Off
Detection
Check the display by disconnecting the
probe.
Waveform and numeric data
should disappear from the
display.
No.
Check Item
Check Procedure
Criteria
10. Cardiac Output (CO)
01
CO Accuracy
Measure the error by inputting 5L/min.
and 15L/min. using the waveform
simulator or CO simulator.
Should be within 0.1~10L/min
±10%.
Should be within 10~20L/min
±15%.
Measure the error by connecting the
waveform simulator or CO simulator,
and inputting the blood temperature of
36
°
C, 37
°
C, 38
°
C.
Should be within ±0.3
°
C for
the blood temperature of
17~45
°
C.
02 Temperature
Accuracy
Measure the error by connecting the
waveform simulator or CO simulator,
and inputting the injectate temperature
of –1
°
C, 35
°
C.
Should be within ±0.5
°
C for
the injectate temperature of
–1~35
°
C.
03 CO
Probe-Off
Detection
Check the display by disconnecting
the probe.
Numeric data should
disappear from the display.
No.
Check Item
Check Procedure
Criteria
11. Carbon Dioxide Concentration (CO
2
)
01 CO
2
Accuracy
Measure the error using the 5%, 10%
calibration gas.
Should be within ±2mmHg for
the range of 0~38mmHg.
Should be within ±5% for the
range of 39~99mmHg.
02 CO
2
Tube-Off
Detection
Check the display by disconnecting
the tube.
Waveform and numeric data
should disappear from the
display.
No.
Check Item
Check Procedure
Criteria
12. External Output (Analog Output)
01 ECG
Output
Sensitivity
Perform ECG input by comprehensive
tester.
Connect the oscilloscope for output
and observe the waveform.
Should be within 1V±10% for
the 1mV input.
Should be within ±0.5V for the
0mV input.
02 BP Output Sensitivity Perform BP input by comprehensive
tester.
Connect the oscilloscope for output
and observe the waveform.
Should be within 1V±10% for
the 100mmHg input.
Should be within ±20mV for
the 0mmHg input.
10
-
16
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...