Periodic Check
Periodic Check (Whole biological information monitoring equipment)
Introduction
The periodic maintenance check is intended to check the medical equipment used daily, to prevent
failures and accidents and to ensure safety and reliability.
As these are general check procedures covering the whole monitoring equipment, perform the
check procedure for the corresponded function of subject equipment.
The check procedures are described for daily check and periodic check. Each check items must
be performed according to the described check procedure.
The consignee can select the check items according to the product quality, frequency of usage, and
maintenance check period. However, electrical safety items must be performed.
For details of the electrical safety check procedure, refer to JIS T 0601-1, IEC 60601-1.
Jig/Measurement Device
The measurement devices used for this periodic check are as follows.
Name Model
Type
Reference
Comprehensive Tester
SG-2000 or equivalent
Chunichi
ECG, RESP, BP, TEMP, CMRR
Waveform Simulator
217A or equivalent
DNI
NEVADA
ECG, RESP, BP, TEMP, CO
Leakage Measurement Jig
3155/3156 or equivalent
HIOKI
Ground Leakage Current, Exterior Leakage Current
(Exterior - Line, Exterior - Ground, Exterior -
Exterior), Patient Leakage Current I/II/III, Patient
Measurement Current
Calibration Gas
GR08081 (5% CO
2
, 21% O
2
)
Oridion
Medical
CO
2
500cc Air Tank
NIBP
AC Resistance Tester
3157-01 or equivalent
HIOKI
Protective Ground Continuity Resistance
Telemetry Receiver
DS-7500 or other telemetry receiver
Fukuda
500cc Air Tank
NIBP
Prepare the following devices as appropriate.
Name Model
Type
Reference
SpO
2
Simulator
Oxitest Plus or equivalent
An approximate check is possible by
measuring healthy subject.
DNI
NEVADA
SpO
2
NIBP Simulator
Cuff Link
An approximate check is possible by
comparing with auscultation on
healthy subject.
DNI
NEVADA
NIBP
Withstand Voltage Tester
3173 or equivalent
HIOKI
Withstand Voltage Test (A-a1, A-a2, B-a)
Reference Power Supply
Equipment
CA-100 or equivalent
Yokogawa
Oscilloscope
Multimeter
10
-
11
Summary of Contents for IB-7300
Page 15: ...Blank Page xii ...
Page 21: ...Blank Page xviii ...
Page 45: ...5 Fix in place with the four screws provided Screw M3x8 4 provided 3 6 ...
Page 71: ...Blank Page 4 22 ...
Page 73: ...Wiring Diagram 5 2 ...
Page 74: ...DC24V Cable 5 3 ...
Page 75: ...Fan Cable 5 4 ...
Page 76: ...AC PWR Cable 5 5 ...
Page 77: ...Ground Terminal Cable 5 6 ...
Page 78: ...Wiring Diagram in the Inlet Unit 5 7 ...
Page 79: ...BlankPage 5 8 ...
Page 85: ...Main Unit Assembly Drawing 410 2905 7 2 ...
Page 86: ...7 3 ...
Page 96: ...External Appearance 1 410 2906 7 13 ...
Page 97: ...External Appearance 2 410 2907 7 14 ...
Page 98: ...Packing Instruction 410 2908 7 15 ...
Page 99: ...Blank Page 7 16 ...
Page 119: ...Blank Page 9 12 ...